beautypg.com

Dataman 520 Series User Manual

Page 65

background image


DATAMAN oscilloscope

User’s Guide

Fig. 5.2.1. – Vertical shift calibration – warm up


The application indicates the time that has elapsed since the oscilloscope start. Once
the ten minutes are reached, the time is displayed in green.

Step 2 – Both oscilloscope inputs should be grounded. In the case where you only
require lower precision, leave the inputs disconnected.

Fig. 5.2.2. – Vertical shift calibration – inputs grounding


Step 3 – The application offers two calibration modes:

Precise calibration – it takes approximately nine minutes and offers the

best precision
Fast

calibration

it

takes approximately three minutes and offers lower

precision. We recommend to perform precise calibration if possible.

- 65 -

This manual is related to the following products: