beautypg.com

Fig. 6.9.2. – status panel – Dataman 520 Series User Manual

Page 72

background image


DATAMAN oscilloscope

User’s Guide


The threshold amplitude is represented by a grey horizontal line in the amplitude part.
You can change the threshold by dragging this line. Phase is displayed only for the
harmonics with amplitude above this threshold.

If the decibels amplitude mode is activated, you can change the amplitude range by
dragging ticks in the left of the amplitude part.

The captured harmonic can be selected by the vertical grey cursor.

You can scroll data using the scroll bar located below the main screen.

The status panel with information about the analyzed signal is located in the left
bottom corner of the main window.


Fig. 6.9.2. – Status panel


If the indicator „Period“ is:

red – the application was not able to find the period automatically

green – period found successfully

grey – the manual selection of the period is activated

The

current

selected

period

is displayed in parentheses.


If the indicator „Timebase“ is:

red, yellow – the timebase is too slow for the analyzed signal, a faster

timebase will produce better results

green – suitable timebase is set

grey – the automatic period selection is deactivated


If the indicator „Vertical shift“ is:

red – the vertical shift is not suitable. If the „Up“ is in parentheses, the

waveform is shifted up, „Down“ – waveform is shifted down.

green – suitable vertical shift is set.

If the indicator „Range“ is:

red – the range is not suitable. If the „Sensitive“ is in parentheses, the

sensitive range is set, „Insensitive“ – the insensitive range is set.

green – suitable range is set.


The information about the harmonic selected by the cursor are displayed in the bottom
of the main window.

- 72 -

This manual is related to the following products: