Product Service 1-800-522-7658
80
Figure 5-10
3.
Pull the tube toward the front of the unit. Remove the access tube cover.
4.
Using an appropriate photometer, connect the unit's upstream line to the access
tube.
Note: Keep the photometer sampling tube as short as possible. It should
never exceed 6 feet maximum.
5.
Turn on the photometer and allow it to operate for a minimum of 5 minutes.
Leave the valve in the “CLEAR” setting.
6.
Check the oil level of the generator – it must be within 1/8” of the level line
when the generator is sitting level.
Calculating the upstream concentration
If you do not want to connect the upstream line of the photometer to the sampling port
located on the supply filter plenum, you must calculate the upstream concentration.
Testing at Labconco has shown the actual concentration varies from the calculated
value. Table 5-2 shows the actual vs. calculated concentrations for the B2 models.
Table 5-2
Type B2 Supply HEPA Filter Leak Test Specifications
Model
30340
30360
Air Displacement (CFM)
445
665
Laskin Nozzles needed
1
1
Theoretical aerosol conc.
(ug/l)
1
30
20
Actual aerosol conc. (ug/l)
1
15
12
1.
The calculated concentration was established by the formula:
Concentration (ug/l) = (# Laskin nozzles @ 20 PSIG x 13,500)/Volume of air displaced
2.
Based on Mineral oil
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132