Labconco PURICARE Procedure Station Models 31260 User Manual
Page 30

Product Service 1-800-522-7658
30
QuickChart – Logic+ PuriCare Procedure Stations
Imperial Units
Model
31241
31142
31261
31162
Type
A2
A1
A2
A1
Cabinet Size (in feet)
4
4
6
6
Sash Opening (inches)
10
12
10
12
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.55
5.55
5.55
Distance between columns
5.21
5.21
5.50
5.50
Inflow Data
Nominal Average Inflow (FPM)
105+/-5
85+/-5
105+/-5
85+/-5
Sash Open Area (Sq. Ft)
3.37
4.04
5.03
6.04
Nominal Avg. Exhaust Vol. (CFM)
354
344
529
514
Avg. Exhaust Vol. Range (CFM)
337-371
323-364
503-553
483-544
Secondary Inflow Data
3
Sash Opening Template
1
3.81
3.81
4.95
4.95
Sensor distance (inches)
2
3.25
3.25
4
4
# of Test points
8
8
12
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (FPM)
354
344
289
281
Avg. Inflow Vel. Range (FPM)
337-371
323-364
275-302
264-297
Correction Factor (CF)
1
1
1.83
1.83
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
337-371
323-364
503-553
483-544
Sash Open Area (Sq. Ft)
3.37
4.04
5.03
6.04
Inflow Velocity Range
(AIV / Sash open area)
100-110
85+/-5
100-110
85+/-5
HEPA Filter Leak Test Data
Air Displacement (CFM)
794
794
1189
1189
Laskin Nozzles needed
2
2
2
2
Theoretical aerosol conc. (ug/l)
3
34
34
23
23
Actual aerosol conc. (ug/l)
4
18
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H
2
O)
<.45
<.45
<.42
<.42
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
353
353
529
529
Performance (Pressure in. H
2
O)
< 0.28”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838501
3838501
3838503
3838503
Motor/Blower Data
Labconco P/N
6
3832201
3832201
3832208
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889201
3889201
3889203
3889203
Canopy Slot Area (F
2
)
.12
.12
.12
.12
Canopy Inflow Range (CFM)
20-100
20-100
20-100
20-100
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal Canopy Slot Volume (CFM)
50
50
50
50
Canopy Vacuum (In. H
2
O)
0.1-0.15
0.1-0.15
0.1-0.15
0.1-0.15
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132