Quickchart – logic+ type a1 - imperial units – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 21

Product Service 1-800-522-7658
21
QuickChart – Logic+ Type A1 - Imperial Units
Model
30132
30142
30152
30162
Type
A1
A1
A1
A1
Cabinet Size (in feet)
3
4
5
6
Sash Opening (inches)
12
12
12
12
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
21 (3x7)
24 (3x8)
30(3x10) 36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.55
5.55
5.55
Distance between columns
4.08
5.21
5.39
5.50
Inflow Data
Nominal Average Inflow (FPM)
85+/-5
85+/-5
85+/-5
85+/-5
Sash Open Area (Sq. Ft)
3.04
4.04
5.04
6.04
Nominal Avg. Exhaust Vol. (CFM)
259
344
428
514
Avg. Exhaust Vol. Range (CFM)
243-274
323-364
403-454
483-544
Secondary Inflow Data
3
Sash Opening Template
1
3+
4+
5+
6+
Sensor distance (inches)
2
4
3.25
3.25
4
# of Test points
6
8
10
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (FPM)
259
344
428
281
Avg. Inflow Vel. Range (FPM)
243-274
323-364
403-454
264-297
Correction Factor (CF)
1
1
1
1.83
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
243-274
323-364
403-454
483-544
Sash Open Area (Sq. Ft)
3.04
4.04
5.04
6.04
Inflow Velocity Range
(AIV / Sash open area)
85+/-5
85+/-5
85+/-5
85+/-5
HEPA Filter Leak Test Data
Air Displacement (CFM)
600
800
1000
1200
Laskin Nozzles needed
1
2
2
2
Theoretical aerosol conc. (ug/l)
3
23
34
27
23
Actual aerosol conc. (ug/l)
4
12
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
36x18x3.06
48x18x3.06
60x18x3.06
72x18x3.06
Performance (CFM)
335
445
555
665
Performance (Pressure in. H
2
O)
< .60”
<.45”
<.45”
<.42”
Labconco P/N
3838400
3838401
3838402
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
18x18x5.88
26x18x5.88
36x18x5.88
48x18x5.88
Performance (CFM)
266
353
441
529
Performance (Pressure in. H
2
O)
< 0.35”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838500
3838501
3838502
3838503
Motor/Blower Data
Labconco P/N
6
3832200
3832201
3832207
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F25T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
Fluor. Lamp Labconco Part #
9721901
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889200
3889201
3889202
3889203
Canopy Slot Area (F
2
)
.12
.12
.12
.12
Canopy Inflow Range (CFM)
20-100
20-100
20-100
20-100
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal Canopy Slot Volume (CFM)
50
50
50
50
Canopy Vacuum (In. H
2
O)
0.1-0.15
0.1-0.15
0.1-0.15
0.1-0.15
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132