Quickchart - logic type a2 bscs – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 26
Product Service 1-800-522-7658
26
QuickChart - Logic Type A2 BSCs
footnotes
Logic Model Identification
The primary serial tag is on the lower outside edge of the right corner post.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet.
The first two digits of the serial number are the year of production; the next
two are the month. The next 5 digits are the sequence of production, and
the letter following the serial number is the revision level of the cabinet.
Downflow Test Specifications
All models are classified as uniform downflow.
All tests performed as described in ANSI/NSF Standard 49:2012.
UV Lamp, IV bar and all other accessories must be removed before
measuring downflow.
Inflow Test Specifications
All tests performed as described in ANSI/NSF Standard 49:2012.
Secondary Inflow Test Specifications
Must use Labconco holder P/N 3836405 to perform this test properly.
1.
Use the appropriate template included in Certifier Kit# 3858403.
2.
Measured from the bottom edge of the probe holder to the center of the
thermal anemometer sensor element.
* - Locate the single row of holes at the front of the grille. Mark the 6
th
hole
from the side wall and subsequent test points every 9 holes until the number of
test points marked equals the width of the cabinet in feet (3-foot cabinet, mark
the first 3 points). Repeat for the opposite side.
HEPA Filter Leak Test Specifications
3.
Based on mineral oil.
4.
The actual aerosol concentration is what was observed during testing.
Aerosol generator should be placed in the left rear corner of the work area, pointing
toward the rear grille. For uncontaminated units, the upstream concentration can be
sampled from the tube located under the work surface
Metric calculation of concentration is:
# of Laskin nozzles @ 138K Pascals x 6.372/Volume of air in m
3
/sec. = Conc. In ug/l
Supply and Exhaust HEPA Filter Specifications
5.
Without gasket
Motor / Blower Specifications
6.
Each motor must be programmed by Labconco for the appropriate width
cabinet. The speed control settings will fluctuate depending on local
temperature and pressure.
UV lamp Specifications
For all models, the UV lamp number is G30T8. The Labconco part number is
1271300.
For all 115/230 VAC models, the ballast assembly is Labconco part number
3829901.
For all 100 VAC models, the ballast assembly is Labconco part number
3830600.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132