Quickchart – logic+ type b2 metric units – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 28

Product Service 1-800-522-7658
28
QuickChart – Logic+ Type B2 Metric Units
Model
30348
30368
Type
B2
B2
Cabinet Size (in m)
1.22
6
Sash Opening (mm)
203
8
Starting Serial #
1303_
1303_
Downflow Data
Nominal Avg. Downflow (m/s)
.279+/.03
55+/-5
grid # of points (rows x columns)
24 (3x8)
36(3x12)
Grid distance from back & sides
152
6.0
Distance between rows (mm)
146
5.75
Distance between columns (mm)
132
5.50
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
105+/-5
Sash Open Area (Sq. m)
.250
.374
Nominal Avg. Exhaust Vol. (m3/s)
.134
.200
Avg. Exhaust Vol. Range (m3/s)
.127-.140
.190-.209
Secondary Inflow Data
3
Sash Opening Template
1
4+
6+
Sensor distance (mm)
2
83
4
# of Test points
8
12
Test point location
*
*
Nominal Avg. Inflow Vel. (m/s)
1.44
1.17
Avg. Inflow Vel. Range (m/s)
1.37-1.50
1.12-1.23
Correction Factor (CF)
.0929
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.127-.140
.190-.209
Sash Open Area (Sq. m)
.250
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
Supply HEPA Leak Test Data
Air Displacement (m3/s)
.210
.471
Laskin Nozzles needed
1
1
Theoretical aerosol conc. (ug/l)
3
30
20
Actual aerosol conc. (ug/l)
4
15
12
Exhaust HEPA Leak Test Data
Air Displacement (m3/s)
.314
.471
Laskin Nozzles needed
2
2
Theoretical aerosol conc. (ug/l)
3
41
27
Actual aerosol conc. (ug/l)
4
25
19
Exhaust Data
DIM Exhaust Volume (m3/s)
5
.314
.471
Traverse Exhaust Volume (m3/s)
6
.382
.575
Differential Pressure (Pa)
7
374
448
Exhaust (Inflow) Alarm Data
Alarm Setpoint (m3/s Inflow)
.071 +/- .03
.105 +/-.03
Supply HEPA Data
Width x Depth x Height (in.)
8
48x18x3.06
72x18x3.06
Performance (CFM)
445
665
Performance (Pressure in. H
2
O)
<.45
<.42
Labconco P/N
3838401
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
8
26x18x8.08
48x18x8.08
Performance (CFM)
720
1100
Performance (Pressure in. H
2
O)
.48+/-.07
.37+/-.07
Labconco P/N
3438501
3438503
Motor/Blower Data
Labconco P/N
6
3832209
3832210
Motor HP
½
½
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132