Labconco PURICARE Procedure Station Models 31260 User Manual
Page 76
Product Service 1-800-522-7658
76
4.
Open the following number of Laskin nozzle(s):
For all 3-foot Logic+ models – 1 nozzle
For all 4-, 5- and 6-foot Logic+ models – 2 nozzles
5.
Ensure the pressure gauge on the generator is reading 20 +/- 1 PSIG.
6.
Scan the downstream side of the HEPA filter by passing the sampling nozzle of
the gun in slightly overlapping strokes over the entire surface of both of the
filters, with the sampling port not more than 1 inch from the surface of the filter
media. Scan the entire periphery of the filters and the gaskets between the filter
frame and the Logic+ frame. Scanning shall be done at a traverse rate of not
more than 2 inches per second.
Note: Operation of the photometer may become erratic when testing the front edge of
the supply filter frame, due to aspiration of room air into the front of the work area. This
problem can be minimized or eliminated by placing the edge of a sheet of rigid plastic or
metal at the edge of the filter gasket when scanning this area, as shown in Figure 5-5.
Figure 5-5
Acceptance
Aerosol penetration shall not exceed 0.01%.
Reassemble the unit by compressing the exhaust filter cover while installing its tabs
back into the exhaust filter bracket slots. Reinstall the diffuser, ensuring that the holes in
the back of the diffuser frame engage the posts on the rear wall of the work area, as
shown in Figure 5-6.
Figure 5-6
Diffuser pins engaging the holes in the rear of
the diffuser.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132