Labconco PURICARE Procedure Station Models 31260 User Manual
Page 168

Product Service 1-800-522-7658
168
PROBLEM
CAUSE
CORRECTIVE ACTION
Fluorescent light
is dim or
flickering
Lamp(s) are
defective
Replace defective lamp(s)
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
UV light not
working
Sash is open
Close sash – UV light will not work
with the sash open.
Lamp is defective
Replace defective lamp.
Lamp wiring is
disconnected.
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections.
UV light is dim or
flickering
Lamp is defective
or is at end of
operating lifetime.
Replace defective or worn out lamp.
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
Airflow Alert
goes off and/or
there is a slight
decrease in filter
life remaining
gauge
HEPA filter loading
The gauge reading steadily
decreases as the Cabinet is used.
Blockage of the
return air slots or
grille
Check all return air slots and grilles
to ensure that they are not blocked
or restricted.
Blockage of the
Ensure that the exhaust outlet is
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132