Electrical Requirements
The biosafety cabinet models have the following electrical requirements:
Model #
Requirements
All 3-& 4-foot Type A ending in -00
115 VAC, 60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -10
100 VAC, 50/60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -20,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All Type B2 ending in -00
115 VAC, 60 Hz, 12 Amps
All Type B2 ending in -10
100 VAC, 50/60 Hz, 12 Amps
All Type B2 ending in -20, 30, 40,
50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All 5-& 6-foot Type A ending in -00
115 VAC, 60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -10
100 VAC, 50/60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -20,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
Note: A dedicated outlet with an appropriate circuit breaker should
be located as close as possible to the right rear side of the cabinet,
at a height even with, or higher than, the top of the bench or stand.
Consult your local electrical codes for properly rated circuit
breakers. For safe operation the dedicated outlet must provide the
protective earthing ground connection to the cabinet.
Note: On 100 and 115 VAC models, both electrical outlets are protected by
a ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.
Do not use any detachable power cord that is not adequately rated
for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
Table 2-1
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132