Quickchart – logic+ type b2 imperial units – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 27

Product Service 1-800-522-7658
27
QuickChart – Logic+ Type B2 Imperial Units
Model
30348
30368
Type
B2
B2
Cabinet Size (in feet)
4
6
Sash Opening (inches)
8
8
Starting Serial #
1303_
1303_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
36(3x12)
Grid distance from back & sides
6.0
6.0
Distance between rows
5.75
5.75
Distance between columns
5.21
5.50
Inflow Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
2.69
4.03
Nominal Avg. Exhaust Vol. (CFM)
283
423
Avg. Exhaust Vol. Range (CFM)
269-296
403-443
Secondary Inflow Data
3
Sash Opening Template
1
3.7
4.95
Sensor distance (inches)
2
3.25
4
# of Test points
8
12
Test point location
*
*
Nominal Avg. Inflow Vel. (FPM)
283
231
Avg. Inflow Vel. Range (FPM)
269-296
220-242
Correction Factor (CF)
1
1.83
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
269-296
403-443
Sash Open Area (Sq. Ft)
2.69
4.03
Inflow Velocity Range
(AIV / Sash open area)
100-110
100-110
Supply HEPA Leak Test Data
Air Displacement (CFM)
445
665
Laskin Nozzles needed
1
1
Theoretical aerosol conc. (ug/l)
3
30
20
Actual aerosol conc. (ug/l)
4
15
12
Exhaust HEPA Leak Test Data
Air Displacement (CFM)
665
998
Laskin Nozzles needed
2
2
Theoretical aerosol conc. (ug/l)
3
41
27
Actual aerosol conc. (ug/l)
4
25
19
Exhaust Data
DIM Exhaust Volume (CFM)
5
665
998
Traverse Exhaust Volume (CFM)
6
810
1218
Differential Pressure (in. H
2
0)
7
1.5
1.8
Exhaust (Inflow) Alarm Data
Alarm Setpoint (CFM Inflow)
150 +/- 5
223 +/- 5
Supply HEPA Data
Width x Depth x Height (in.)
8
48x18x3.06
72x18x3.06
Performance (CFM)
445
665
Performance (Pressure in. H
2
O)
<.45
<.42
Labconco P/N
3838401
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
8
26x18x8.08
48x18x8.08
Performance (CFM)
720
1100
Performance (Pressure in. H
2
O)
.48+/-.07
.37+/-.07
Labconco P/N
3438501
3438503
Motor/Blower Data
Labconco P/N
6
3832209
3832210
Motor HP
½
½
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132