Labconco PURICARE Procedure Station Models 31260 User Manual
Page 113
Product Service 1-800-522-7658
113
3.
Remove the last two screws. Gently rock the cover, slide it to the left or
right, and swing it out and away from the unit.
4.
Use 1/2- and 7/16-inch sockets or wrenches to tighten the three plenum lift
bolts located as noted in Figure 6-27. This will lift the plenum off of the
supply HEPA filter. Remove the Supply HEPA filter retainers on the front of
the plenum, if present.
5.
Slide the supply HEPA filter straight out the unit.
6.
With the HEPA filters removed, inspect the filter shelf for damage.
Figure 6-27
1/2” Plenum Lift Bolt
7/16” Plenum Lift Bolt
Supply HEPA filter retainers (3)
To Install a Supply HEPA Filter:
1.
Install a new supply HEPA filter by pushing it straight into the cabinet,
ensuring that it is correctly oriented, and that the filter is centered on the
filter shelf.
2.
Use the 7/16- and 1/2-inch sockets or wrenches to sequentially loosen the
three plenum lift bolts located as noted in Figure 6-27. When properly
positioned, the plenum box will be aligned with the supply HEPA filter frame.
Continue to loosen the bolts until the plenum contacts the HEPA filter and
has slightly compressed the filter gasket.
NOTE: The lift bolts should only be tightened enough to ensure a proper seal
of the supply HEPA filter. At maximum tightness, the supply filter's gasket
should be compressed to less than 50% of its original thickness. NEVER
tighten the locking bolts beyond this setting.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132