Labconco PURICARE Procedure Station Models 31260 User Manual
Page 131
Product Service 1-800-522-7658
131
6.
Next, connect the meters + DC voltage lead to pin 6 (yellow wire) and verify
there is ± 24 Volt DC ± 5 Volts DC, as shown in Figure 7-10. The presence of
24 Volt DC on this yellow wire tells the motor to start.
Figure 7-10
Measuring 24 VDC voltage
7.
Next, move the meter’s + DC voltage lead to pin 5 (purple wire) [the meter’s lead
is still connected to pin 8 (brown wire)]. Note the DC voltage, since the motor’s
speed is determined by the pulse width modulated signal on this contact. A
voltage noted in step 6 signifies 100% pulse width duty cycle, while a DC voltage
of 0 volts would indicate a 0% PWM level. Depending on the particular ECM
motor, the motor speeds for the 0% PWM levels and 100% PWM levels will differ
but the 0% level represents the “night setback” blower speed and the 100% level
indicates the max speed ever required by the motor under total filter loading. The
actual DC voltage measured will be between 0 V and 24 Volts DC, and that voltage
indicates roughly the % duty cycle being sent to the motor at this time. For
example:
A total voltage of 22 VDC, and a signal voltage of 15 VDC would indicate 15 VDC
measured/22V DC @ 100% = 68% duty cycle.
Another example, would be 18 VDC/22 DC @ 100%, would correspond to an 81%
PWM signal.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132