To Remove the Supply HEPA Filter:
1.
With the exhaust filter removed, use a 1/2-inch socket or wrench to tighten the
three plenum lift bolts located as noted in Figure 6-22. This will lift the plenum off of
the supply HEPA filter. Remove the Supply HEPA filter retainers on the front of the
plenum.
2.
Slide the supply HEPA filter straight out the unit.
3.
With the HEPA filters removed, inspect the filter shelf and the exhaust filter tray for
damage.
Figure 6-22
Plenum Lift Bolts (3)
Note: The Blower has been
removed for clarity only.
Supply HEPA filter retainers (3)
To Install a Supply HEPA Filter:
1.
Install a new supply HEPA filter by pushing it straight into the cabinet, ensuring that
it is correctly oriented, and that the filter is centered on the filter shelf.
2.
Use a 1/2-inch socket or wrench to sequentially loosen the three plenum lift bolts
located as noted in Figure 6-17. When properly positioned, the plenum box will be
aligned with the supply HEPA filter frame. Continue to loosen the bolts until the
plenum contacts the HEPA filter and has slightly compressed the filter gasket.
NOTE: The lift bolts should only be tightened enough to ensure a proper seal
of the supply HEPA filter. At maximum tightness, the supply filter's gasket
should be compressed to less than 50% of its original thickness. NEVER
tighten the locking bolts beyond this setting.
To Install an Exhaust HEPA Filter:
1.
On 3-foot wide models, reinstall the exhaust filter pan, ensuring the filter stop flange
is oriented toward the rear of the cabinet.
2.
On 3-foot wide models, thread all four sealed bolts into the exhaust filter pan.
3.
Slide the new exhaust HEPA filter all the way into the pan.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132