Type a exhaust hepa filter test – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 77

Product Service 1-800-522-7658
77
Type A Exhaust HEPA Filter Test
Opening the Exhaust HEPA Filter Cover
If there is sufficient overhead clearance (18 or more inches) –
1.
Loosen the two Phillips screws on the front sides of the filter cover, as shown in
Figure 5-7. Tilt the cover back access the downstream side of the filter, as shown
in Figure 5-8.
Figure 5-7
Loosen these screws to tilt the
cover back.
Remove these acorn nuts to
remove the cover.
If there is insufficient overhead clearance –
1.
Loosen the two Phillips screws on the front sides of the filter cover, and remove
the four acorn nuts and washers. Carefully lift the cover off of the top of the
cabinet.
Note: exercise caution when removing or installing the exhaust filter cover,
to prevent damaging the Exhaust HEPA filter.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132