Labconco PURICARE Procedure Station Models 31260 User Manual
Page 74
Product Service 1-800-522-7658
74
If the Logic+ is Uncontaminated
(Establishing an Upstream Concentration)
The Purifier utilizes a unique aerosol access tube for sampling the upstream
concentration of aerosol. If after reading these instructions, you have further questions,
please contact Labconco's Product Service Department.
Note: The access tube contains contaminated air under positive pressure. DO
NOT access this line until the cabinet has been decontaminated.
1.
Remove the work surface by lifting the handles on either end. Pull the work
surface straight out of the Logic+.
Locate the access tube on right side of the rear plenum wall, behind through the towel
catch. As shown in Figure 5-2.
Figure 5-2
2.
Pull the tube toward the front of the unit. Remove the access tube cover.
3.
Using an appropriate photometer, connect the unit's upstream line to the access
tube.
Note: Keep the photometer sampling tube as short as possible. It should
never exceed 6 feet maximum.
4.
Turn on the photometer and allow it to operate for a minimum of 5 minutes.
Leave the valve in the “CLEAR” setting.
5.
Check the oil level of the generator – it must be within 1/8” of the level line
when the generator is sitting level.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132