Circuit Breaker Specifications, 230 VAC
Cabinet Width (feet)
3
4
5
6
Motor/Light Breaker Rating,
Amps
4
4
5
5
Motor/Light Breaker, Labconco
Part #
1327209 1327209
1327204 1327204
Motor/Light Breaker,
Mechanical Products Part #
1600-062-
040-007
1600-062-
040-007
1600-062-
050-007
1600-062-
050-007
Outlet Breaker Rating, Amps
3
3
3
3
Outlet Breaker, Labconco
Part #
1327200 1327200
1327200
1327200
Outlet Breaker, Mechanical
Products Part #
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
Testing a circuit breaker
Note: If the breaker is tripped (white button is extended), press the reset button
of the breaker. If it does not reset smoothly, the breaker has failed and must be
replaced.
1.
With the breaker reset, unplug the Logic+, either at the wall, or at the power
cord inlet at the back of the electronics module, as shown in Figure 9-2.
2.
Remove the three 7/16-inch acorn nuts and lock washers that secure the
module.
3.
Tip the module on its side to access the breakers.
4.
Using a continuity meter, check the continuity (resistance) across the
terminals of the breaker being tested.
5.
The resistance across the breaker should not exceed the resistance noted
when the meter probes are connected to each other.
Replacing a circuit breaker
1.
Unplug the Logic+, either at the wall, or at the power cord inlet at the back
of the electronics module, as shown in Figure 7-2.
2.
Remove the two 7/16-inch acorn nuts and lock washers that secure the
module.
3.
Tip the module on its side to access the breakers.
4.
Disconnect the wires connected to the breaker terminals.
5.
Using a pair of pliers, loosen the chrome ring nut outside of the reset button.
Remove the ring nut, and pull the breaker out of the electronics module.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132