Labconco PURICARE Procedure Station Models 31260 User Manual
Page 167

Product Service 1-800-522-7658
167
PROBLEM
CAUSE
CORRECTIVE ACTION
Blower runs
slower than
normal; speed
control has little
effect on blower
speed. Blower may
take more than 5
seconds to start.
Motor signal wires
damaged or
disconnected
Measure motor output signal
voltage. Inspect signal wiring and
connections
Power Supply
Board
misconnected or
defective
Test output voltages on the Power
Supply board.
Motor jumper wire
is damaged/
disconnected
(115/100 Volt
models only)
Inspect grey wire #31 on 15 pin
connector for continuity.
Blower motor is
defective
Replace blower motor.
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections.
Fluorescent light
not working
Sash is closed
Open sash – Fluorescent lights will
not work with the sash closed.
Lamp(s) are
defective
Replace defective lamp(s)
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballasts
Replace lamp ballasts.
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132