Quickchart – logic+ type a1 - metric – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 22

Product Service 1-800-522-7658
22
QuickChart – Logic+ Type A1 - Metric
Model
30132
30142
30152
30162
Type
A1
A1
A1
A1
Cabinet Size (in meters)
.91
1.22
1.52
1.83
Sash Opening (inches)
305
305
305
305
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (m/s)
.279 +/-.03
.279 +/-.03
.279 +/-.03
.279 +/-.03
grid # of points (rows x columns)
21 (3x7)
24 (3x8)
30(3x10)
36(3x12)
Grid distance from back & sides
152
152
152
152
Distance between rows (mm)
141
141
141
141
Distance between columns (mm)
104
132
137
140
Inflow Data
Nominal Average Inflow (m/s)
.432 +/-.03
.432 +/-.03
.432 +/-.03
.432 +/-.03
Sash Open Area (Sq. m)
.283
.375
.468
.561
Nominal Avg. Exhaust Vol. (m3/s)
.122
.162
.202
.242
Avg. Exhaust Vol. Range (m3/s)
.115-.129
.153-.172
.190-.214
.228-.257
Secondary Inflow Data
3
Sash Opening Template
1
3+
4+
5+
6+
Sensor distance (mm)
2
102
83
83
102
# of Test points
6
8
10
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (m/s)
1.32
1.75
2.17
1.43
Avg. Inflow Vel. Range (m/s)
1.23-1.39
1.64-1.85
2.05-2.31
1.34-1.51
Correction Factor (CF)
.0929
.0929
.0929
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.115-.129
.153-.172
.190-.214
.228-.257
Sash Open Area (Sq. m)
.283
.375
.468
.561
Inflow Velocity Range
(AIV / Sash open area)
.432 +/-.03
.432 +/-.03
.432 +/-.03
.432 +/-.03
HEPA Filter Leak Test Data
Air Displacement (m/s)
.283
.378
.472
.566
Laskin Nozzles needed
1
2
2
2
Theoretical aerosol conc. (ug/l)
3
23
34
27
23
Actual aerosol conc. (ug/l)
4
12
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
36x18x3.06
48x18x3.06
60x18x3.06
72x18x3.06
Performance (CFM)
335
445
555
665
Performance (Pressure in. H
2
O)
< .60”
<.45”
<.45”
<.42”
Labconco P/N
3838400
3838401
3838402
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
18x18x5.88
26x18x5.88
36x18x5.88
48x18x5.88
Performance (CFM)
266
353
441
529
Performance (Pressure in. H
2
O)
< 0.35”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838500
3838501
3838502
3838503
Motor/Blower Data
Labconco P/N
6
3832200
3832201
3832207
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F25T8 TL741 F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721901
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889200
3889201
3889202
3889203
Canopy Slot Area (m
2
)
.0111
.0111
.0111
.0111
Canopy Inflow Range (m/s)
.010-.047
.010-.047
.010-.047
.010-.047
Nominal Canopy Slot Velocity (m/s)
1.22-1.32
1.22-1.32
1.22-1.32
1.22-1.32
Nominal Canopy Slot Volume (m3/s)
.235
.235
.235
.235
Canopy Vacuum (Pa)
25-37
25-37
25-37
25-37
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132