Type b2 prefilter replacement, Type b2 hepa filter replacement – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 111
Product Service 1-800-522-7658
111
Type B2 Prefilter Replacement
NOTE: The prefilter is not washable; it must be replaced when it becomes
dirty.
1.
Shut off the cabinet blower.
2.
Using an appropriate step or ladder, remove the two wing nuts that secure the
prefilter, located next to the electronics module, as shown in Figure 6-24.
3.
Remove and discard the prefilter, and replace it with a new one. Retighten the two
wing nuts that secure the prefilter.
Figure 6-24
Prefilter wing nuts
Type B2 HEPA Filter Replacement
CAUTION: The cabinet must be properly decontaminated before servicing the
exhaust HEPA filter. Following replacement of either HEPA filter, the cabinet
must be recertified.
CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate
lifting techniques to remove and handle it. Use two people to remove the
larger filters if possible.
1.
Unplug the cabinet.
2.
Remove the front dress panel as shown in Figure 6-1.
To Remove the Exhaust HEPA Filter:
3.
Loosen all of the Phillips screws on the exhaust filter cover.
4.
Loosen the four sealed exhaust filter bolts on the top of the cabinet hull, as
shown in Figure 6-25. DO NOT remove the bolts; only loosen them until the
exhaust HEPA filter is free of the filter frame.
5.
Pull the exhaust HEPA filter straight out of the cabinet.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132