Quickstart – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 11

Product Service 1-800-522-7658
11
Quickstart
How do I recalibrate the
blower speed
filter life gauge
airflow sensor reading
of the Logic+? - Continued
11.
If this is an initial certification or a recertification after HEPA filter
replacement, select
Reset Filter Gauge to 100%
. If this is an
annual recertification, select
Leave Gauge unchanged
. If you
want to set the filter gauge at a preset value, select
Set Gauge
to new setting
. Then press
OK/Mute
.
12.
The Type A units with airflow sensor will ask you to now
calibrate the average downflow to match the value you obtained
when you measured it. After the downflow is calibrated, and you
press
OK/Mute
, the Logic+ will power down, and then restart
with the new settings.
13.
If the unit is a Type A without an airflow sensor, after the filter
gauge option is selected and you press
OK/Mute
, the Logic+
will power down, and then restart with the new settings. If the
Unit is a Type B2 or a Type A with an airflow sensor, you will
then be asked to calibrate the inflow sensor to match the value
you obtained when you measured the inflow. After the inflow is
calibrated, and you press
OK/Mute
, the Logic+ will power
down, and then restart with the new settings.
14.
For the Type B2, the unit will now ask you to reduce the
total
exhaust volume to 80% of the nominal value. When you have
done this, press
OK/Mute
.
This step is critical-it sets the alarm
point for the exhaust flow. If you press OK/Mute before reducing
the flow to 80%, the Logic+ will assume the alarm setpoint is
equal to the nominal airflow, and the unit will return an Alarm-
Inflow Screen.
The Logic+ will power down, and then restart
with the new settings. Be sure to reset the exhaust damper to its
nominal setting.
Calibration
Filter Gauge Adjustment
or
to change filter gauge,
OK to select, MENU to finish
Reset Filter Gauge to 100%
Leave Gauge unchanged
Set Gauge to new setting
Reduce the exhaust flow to
80% of nominal volume. Press
OK when you are ready to fix
the setpoint.
Calibration
Inflow Alarm Setpoint
OK to select, MENU to finish
The Inflow displayed
value is: 102 FPM
Calibration
Inflow Display
or
to change display value,
UV Light to reset to 105 FPM,
OK to select, MENU to finish
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132