Preparing the logic+ for gaseous sterilization – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 120

Product Service 1-800-522-7658
120
Preparing the Logic+ for Gaseous
Sterilization
Note: This section only reviews the steps required for preparing the Logic+
for Gas Sterilization. Thoroughly understand the sterilization procedures and
protocols supplied by the manufacturer of the sterilizing system before
attempting this operation.
1.
Unplug the Logic+ from electrical power.
2.
Thoroughly surface decontaminate the working area of the cabinet.
In order to get a complete seal on the front of the cabinet, do the following:
1.
Remove the work surface after it has been surface decontaminated.
2.
Remove the towel catch, and locate the Sampling/Decontamination Tube, as
shown in Figure 6-32.
3.
Remove the front grille.
4.
Place the work surface and the grille into the work area for decontamination.
5.
Remove the front dress panel.
6.
Remove the sash stop bracket, located near the top of the blower plenum cover.
7.
Lift the sash up until the bottom edge of the sash clears the bottom edge of the
Supply HEPA Filter shelf.
8.
Using flexible plastic sheet 5-8 mil thick and duct tape seal the perimeter of the
front opening. If you wish to connect the sterilizer’s discharge tube to the
Logic+’s Sampling/Decontamination Tube, do this now, and properly seal the
discharge tube to the plastic sheet.
9.
Open the exhaust filter cover.
Note: DO NOT attempt to remove the brackets that secure the exhaust filter cover;
they are needed to maintain rigidity on the top of the chassis.
10.
Using flexible plastic sheet and duct tape seal the Exhaust filter opening.
11.
The Logic+ is now prepared for gaseous decontamination.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132