60
Alarm Screens
Sash is too high
The sash is open too far for safe operation.
Airflow Alert
The airflow patterns in the cabinet have changed,
resulting in a sudden change in the motor speed.
This is most likely due to a blockage of the grille or
the exhaust filter outlet. It may also be caused by
removal of the work surface while the cabinet is in
operation.
Airflow Alarm
If equipped with the optional airflow sensor and
the inflow or downflow velocities are excessively
high or low, this alarm will be displayed.
Canopy Alarm
If equipped with the optional canopy connection,
and there is insufficient exhaust system airflow,
this alarm will be displayed. If this alarm is
displayed, the Logic+ blower will need to be
turned off, and then back on again to reset the
alarm.
System Error
The motor and display circuit board are not
communicating properly. DO NOT USE THE
CABINET UNTIL THE PROBLEM HAS BEEN
CORRECTED.
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132