Motor/blower replacement – Labconco PURICARE Procedure Station Models 31260 User Manual
Page 114
Product Service 1-800-522-7658
114
Motor/Blower Replacement
Caution: The cabinet must be properly decontaminated before servicing the
motor/blower.
The cabinet blower motor's bearings are sealed and require no lubrication.
DO NOT attempt to lubricate them.
To replace the Motor/Blower:
1.
Unplug the cabinet.
2.
Remove the front dress panel as shown in Figure 6-1.
3.
Loosen all of the Phillips screws on the blower chamber cover. Remove all of the
screws except two screws on the top center of the panel. This will support the
panel.
4.
Release the damper seal by pushing the outside edge of the seal towards the
bolt and inside the blower plenum cover, as shown in Figures 6-14 and 6-15.
5.
Remove the last two screws. Gently rock the cover, slide it to the left or right,
and swing it out and away from the unit.
NOTE: During disassembly, the blower plenum cover can rest on the central sash
support, as shown in Figure 6-16, to allow for repositioning of the panel or your grip on
it.
6.
Disconnect both motor connectors from the motor. The 5-pin power connector,
located closest to the blower, is released by pressing the locking ears on either
side of the connector. The 16-pin data connector, located furthest from the
blower, is released by pressing the single locking ear located on the inside center
surface of the connector body. The connectors are shown in Figure 6-28.
Figure 6-28
ECM Power Connector Release
Tabs
ECM Data Connector Release Tab
- PURICARE Procedure Station Models 31240 PURICARE Procedure Station Models 31162 PURICARE Procedure Station Models 31142 Type B2 Biosafety Cabinet Models 32368 Type B2 Biosafety Cabinet Models 32348 Type B2 Biosafety Cabinet Models 30368 Type B2 Biosafety Cabinet Models 30348 Type A2 Biosafety Cabinet Models 34269 Type A2 Biosafety Cabinet Models 34259 Type A2 Biosafety Cabinet Models 34249 Type A2 Biosafety Cabinet Models 34239 Type A2 Biosafety Cabinet Models 33269 Type A2 Biosafety Cabinet Models 33259 Type A2 Biosafety Cabinet Models 33249 Type A2 Biosafety Cabinet Models 33239 Type A2 Biosafety Cabinet Models 32269 Type A2 Biosafety Cabinet Models 32259 Type A2 Biosafety Cabinet Models 32249 Type A2 Biosafety Cabinet Models 32239 Type A2 Biosafety Cabinet Models 30268 Type A2 Biosafety Cabinet Models 30258 Type A2 Biosafety Cabinet Models 30248 Type A2 Biosafety Cabinet Models 30238 Type A2 Biosafety Cabinet Models 30261 Type A2 Biosafety Cabinet Models 30251 Type A2 Biosafety Cabinet Models 30241 Type A2 Biosafety Cabinet Models 30231 Type A1 Biosafety Cabinet Models 30162 Type A1 Biosafety Cabinet Models 30152 Type A1 Biosafety Cabinet Models 30142 Type A1 Biosafety Cabinet Models 30132