Checkline FH Series User Manual
Page 59

© ElektroPhysik MiniTest 7200 FH / MiniTest 7400 FH
59
15.2 Detection of Error sources (INF flag)
Measurement may be influenced by static or alternating magnetic fields and by metallic objects
near the sensor.
INF flag
INFdis-
appeared
If no target ball is placed on the sensor, the “INF” flag
appears on display. That means that the probe is in
the infinity range.
Make sure that INF is shown so that correct
readings can be taken.
If the sensor is subject to some interference (e.g.
through electro-motors, transformers, potential
distribution or if a steel plate is near the sensor) the
INF flag disappears.
Try to move the sensor a little bit in order to locate the
source of interference. If you have detected the error
source, move the sensor away from it. As soon as the
“INF” flag appears on display, the sensor is out of the
error source range and the gauge is ready again to
take correct measurements.
To be sure, take double the distance form the error
source.
In the default measure mode using a target ball, it is
not of importance if the “INF”-flag disappears.