15. Appendix

15.1 Troubleshooting

Error message

Problem

Remedy

“Please check clock settings“

If you insert batteries at initial use

or if the gauge has been cut from

power supply for more than 5

minutes

Reset clock (s. section 10.5.3)

“Low Bat !!!”

Though you can continue to

measure, it is recommended to

change batteries. If battery

voltage is below minimum, the

gauge will switch off

automatically.

Change batteries. Dispose

batteries according to your local

regulations.

“Please hold probe in air(infinite

range) !! “

(Briefly appears on display at

switch on)

Please check

a) whether you have removed the

shielding tube or protection cap

b) whether a target ball is

positioned on the sensor tip

c) whether the sensor tip is in

contact with a steel part

See also section 6.1

a) Remove shielding tube or

protection cap

b) Remove target ball from sensor

tip

c) Enlarge distance between

sensor head and steel part

Now the message on display

should disappear and the

sensor is ready for

measurement.

“Confirm sensor in infinite range”

a) a target ball is placed on the

sensor tip or the sensor is near an

ferritic object or magnetic field.

b) Infinite adjustment required.

See also section 6.1

The message does not

disappear, and the following

message appears after about

10 sec:

“Make sure there is no target ball

near the sensor.

Please keep the sensor away

from metallic objects or magnetic

fields.”

If this message persists, press