Target ball: (adjusted ball size)

measurement.

“Full memory capacity !”

Probable cause:

You have reached maximum

number of storable readings.

You have reached maximum

number of entries in the data

base (MiniTest 7400 only)

Delete readings, batches or

directories in the database which

you do not need any more.

“Memory optimization in

progress”

Message appears briefly on

display.

Automatic procedure. No further

action required.

“Check sensor connection !”

Possible cause:

the sensor has been

removed

loose probe connection

damaged cable

defective sensor

Please check the options. If the

sensor is defective, please

change it or contact after sales

service.

“No sensor connected ! “

“Changes not possible !”

Sensor connecting error.

Possible cause:

no sensor connected

loose probe connection

damaged cable

defective sensor

Please check the options. If the

sensor is defective, please

change it or contact after sales

service.

“Sensor failure “”

Please contact after-sales

service.

The following errors may be remedied by a total reset (see section 11.1)

-

keyboard failure

-

no measuring admitted

-

illogical readings

If the gauge cannot be switched off via the ON/OFF button, remove and reinsert batteries.