Note:
Continuous print mode may be used in conjunction with Timed Stop to
print data throughout the time period. Printing will occur only during
the test.
S
TIMED AVERAGE
The DV-III Ultra will collect a specified number of data points over a period of time and present
the average reading (arithmetic average). Time, speed and the number of data points are entered
using the number keys and
OPTIONS/TAB
key. The test will begin immediately upon the press of
the
ENTER
key. Non-averaged data will be displayed during the test.
The test will end when the specified time interval has elapsed. The averaged data point will be
shown as displayed in Figure V-7. The
SELECT DISP
key can be used to view the average value of
all measurement parameters. The
key can be used to send a single averaged dataline to the
printer.
DATA AVERAGING
300 READINGS WERE
AVG'D OVER 02:30 M:S
AVERAGE TORQUE=90.3
Figure V-7
Note:
Printing during a Timed Average test will show non-averaged data.
Averaged data is available only at the conclusion of the test. Printing the
averaged data must be done with the
key at the end of the test.