VISCOMETER SETUP

PRESS ENTER TO START

Figure II-9

Press the

ENTER

key. The screen in Figure II-10 appears. Press the "3" key for NO.

PRINT PARAM RAM?

1 = YES

3 = NO

Figure II-10

The display will now ask “SPECIAL SPINDLE?” as shown in Figure II-11. Press the “1” key

for YES.

SPECIAL SPINDLE?

1 = YES

3 = NO

Figure II-11

The display will ask “ADD A SPINDLE?” Press the “1” key for YES.

ADD A SPINDLE?

1 = YES

3 = NO

Figure II-12

The spindle number and name will be flashing. Use the “2” key for scrolling downward (ê) and

the “8” key for scrolling upward (è). The spindle name can be changed to any (2) letters from A

to Z.
Use the “tab” key to move left and right through the menu. Press the “enter” key to accept.

SPECIAL SPINDLE #1
NAME = D AA E
SMC =
SRC =

Figure II-13

Check the spindle constants to be entered. You must have the SMC (spindle multiplier constant)

and the SRC (the shear rate constant). Enter the first set of numbers. Press the “enter” key to go

to the next line (example: spindle AA = SMC = 1.0 SRC = 2.0).
If you are adding additional spindles, press the “enter” key. You can enter up to 5 special

spindles.