Brookfield TA-DGA User Manual
Ta-dga – dual grip assembly

11 Commerce Boulevard, Middleboro, MA 02346 USATel: 800-628-8139 or 508-946-6200 • Fax: 508-946-6262 
Email: [email protected] 
www.brookfieldengineering.com
M11-375-TADGA
Installation Instructions for Texture
Accessory Part Number: TA-DGA
TA-DGA – Dual Grip Assembly
The Dual Grips are used to pull a sample apart or peel the 
backing from its substrate in a 180-degree peel test. The 
sample is secured into the grips after they have been brought 
close together. When the test is run, the sample is pulled 
apart and the distance and load force are recorded. 
Fixture Base Table (TA-BT-KIT)
No base table is used with these grips.
Install the Dual Grips
The lower grip attaches directly to the slot in the base of the 
CT3 by using the two T-bolts supplied. Leave the locking 
T-bolts loose until the alignment is complete. The upper 
grip attaches directly to the M6 thread on the probe shaft. 
Screw the upper grip fully into the probe shaft, aligning the 
grip faces with those of the lower grip. Lock into position. 
 
Alignment of the Dual Grips in Stand-Alone Mode
The easiest way to align the lower grip under the upper grip 
in stand-alone mode is by using the Tension test. Rotate the 
SELECT/SCROLL knob until Tension test appears on the 
display. After depressing the start button two times, the 
SELECT/SCROLL knob can be used to lower the upper 
grip. Depressing and holding the Select/Scroll knob will 
continuously lower the probe, and rotating the SELECT/SCROLL knob will lower the probe 1 mm for each 
click of the knob. Bring the probe down 100 mm (this position will be shown in the CT3 display), or until it 
almost touches the lower grip. Clamp a flat, rigid sheet of paper or thin cardboard into the center of the 
grip. Align the lower grip directly under the upper grip using the clamped sheet as a guide. The grips 
should now be properly aligned. Lock the lower grip by tightening the T-bolts. 
 
Use of the Dual Grip Accessory 
Once the Dual Grips are aligned, tests can be run in either stand-alone mode or with Texture-Pro CT 
Software. 
 
In stand-alone mode, choose the Tension test. Set a Trigger value appropriate for the load cell in your 
instrument, usually 20 to 50 grams. The maximum deformation distance possible in the CT3 is 101 mm if 
the test starts with the top grip in its lowest position. The position of the grip is always displayed on the 
screen. The deformation distance for the test must be no more than this displayed value when the Trigger is 
