beautypg.com

Brookfield DV-III Ultra Rheometer User Manual

Page 87

background image

Brookfield Engineering Laboratories, Inc.

Page 87

Manual No. M98-211-E0912

IV.8.6.3 Test Results

Test Passed

When a yield test runs successfully, the resultant Yield Stress (Pa), the % Torque value at the time

of the yield and the Temperature at yield are printed.

Test (#5: Sample Tst) Complete

Yield Stress (Pa) = 726.36 % Torque @ Yield = 73.56 Temperature = 25.5C

In addition, the results are shown on the display as follows:

#5: Sample Tst S:71

TEST COMPLETE

%=73.56 TEMP=025.5
YIELD=0726.4 Pa

Figure IV-84

Test Failed

Some of the reasons a yield test may fail are as follows.

Under-Range Condition

Only yield measurements in the % Torque range from 10% to 100% will be accepted. If a Yield

Stress is reached before the % Torque value is 10%, an Under-Range error causes the test to end.

The following is sent to an attached printer:

Test (#5: Sample Tst) Complete

Test Failed = Under-range

The following screen is displayed:

#5: Sample Tst S:71

YIELD ERROR

UNDER-RANGE

PRESS ESC OR ENTER

Figure IV-85

Under-Range Error

The yield stress (Torque Reduction Value) is reached before the absolute % torque reaches 10%.

When a test is started, once the absolute torque is greater than zero (0), the second, third, and fourth

Delta Torques (also known as the Base Increment Torque) are averaged. The first three (3) Delta

Torques after 9% absolute torque is reached are also averaged. An under-range error occurs if the

second average falls below the first average by 15% or more.

Over-Range Condition

Only yield measurements in the % Torque range from 10% to 100% will be accepted. If Yield

Stress is not reached before the measured % Torque value reaches 100%, an Over-Range error

causes the test to end.

The following is sent to an attached printer:

Test (#5: Sample Tst) Complete

Test Failed = Over-range