Brookfield DV-III Ultra Rheometer User Manual
Page 47
Brookfield Engineering Laboratories, Inc.
Page 47
Manual No. M98-211-E0912
Command
Command Description
Parameter
Comments
SSP
Set spindle
31
set to a number 31 spindle
SSN
Set viscometer speed
100.0
run at 100 RPM
SPI
Set print interval
00:05
begin printing data at 5 second
intervals
WPT
Wait for % torque
85.0
wait until % torque = 85; a curing
cycle
Program 4: Spring relax
Command
Command Description
Parameter
Comments
WKY
Wait for a key press
Wind to 100%
tell operator to wind spindle
until 100% torque is reached
WPT
Wait for % torque
100.0
wait until 100% torque is reached
WKY
Wait for a key press
Release spindle
tell operator to release the spindle
SPI
Set print interval
00:01
begin printing data at 1 second
intervals
WPT
Wait for % torque
0.0
wait for spindle to completely
unwind to 0% torque
Program 5: Variable speed
Command
Command Description
Parameter
Comments
SSN
Set viscometer speed
50
run at 50 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
10
run at 10 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
20
run at 20 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
50
run at 50 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
IV.3.2
Running B.E.V.I.S Programs in Standalone Mode
The B.E.V.I.S. program menu is accessed by pressing the PROG key and selecting number 2,
2 = B.E.V.I.S. The loading and execution of B.E.V.I.S. programs are described in the following
sections.