Brookfield DV-III Ultra Rheometer User Manual
Page 102

Brookfield Engineering Laboratories, Inc.
Page 102
Manual No. M98-211-E0912
Thermosel
Spindle
Shear
Rate
(sec-
1
)
Viscosity (cP)
LVDV-III Ultra
RVDV-III Ultra
HADV-III Ultra
HBDV-III Ultra
18
0-264 1.3 -
30,000
12.8 -
320,000
25.6 -
640,000
103 - 2,560,000
21
0-186 1.9 -
46,865
20.0 -
500,000
40.0 - 1,000,000
160 - 4,000,000
27
0-68 9.4 - 234,325 100.0 - 2,500,000 200.0 - 5,000,000
800 - 20,000,000
28
0-56 18.8 - 468,650 200.0 - 5,000,000 400.0 - 10,000,000 1,600 - 40,000,000
29
0-50 37.5 - 937,300 400.0 - 10,000,000 800.0 - 20,000,000 3,200 - 80,000,000
31
0-68 12.0 –
300,000 128.0 - 3,200,000 256.0 - 6,400,000 1,024 - 25,600,000
34
0-56 24.0 - 600,000 256.0 - 6,400,000 512.0 - 12,800,000 2,048 - 51,200,000
81
0-258 1.4 -
10,000
15.0 -
10,000
30.0 -
10,000
120 -
10,000
UL Adapter
UL Spindle
Shear Rate
(sec-
1
)
Viscosity (cP)
LVDV-III Ultra
RVDV-III Ultra
HADV-III Ultra
HBDV-III Ultra
YULA-15 or
15Z
0 - 122.4
1.0 - 2,000
3.2 - 2,000
6.4 - 2,000
25.6 - 2,000
Note: LV DV-III Ultra - Minimum viscosity, 1.0 cP at 60RPM, 10% of full scale range.
Note: RV,HA,HB-III - Minimum viscosity; 3 cP, 6 cP and 24 cP, at 230.9 RPM, 10% of full scale
range.
DIN Adapter
DDA
Spindle
Shear Rate
(sec-
1
)
Viscosity (cP)
LVDV-III Ultra
RVDV-III Ultra
HADV-III Ultra
HBDV-III Ultra
85
0 - 258
0.6 - 5,000
6.1 - 5,000
12.2 - 5,000
48.8 - 5,000
86
0 - 258
1.8 - 10,000
18.2 - 10,000
36.5 - 10,000
146 - 10,000
87
0 - 258
5.7 - 50,000
61.0 - 50,000
121 - 50,000
485 - 50,000