Brookfield DV-III Ultra Rheometer User Manual
Page 81
Brookfield Engineering Laboratories, Inc.
Page 81
Manual No. M98-211-E0912
Figure IV-68: Stress vs. Strain Graphs for Various Food Products
IV.8.6
Running Yield Test Programs in Stand-Alone Mode
This section explains how to use the DV-III Ultra Rheometer in standalone mode. There are no test
programs stored in instrument memory when the DV-III Ultra is shipped from Brookfield. Therefore,
you will need to review the preceding section on the EZ-Yield
TM
Software which explains how to
create test programs. Once a test program has been downloaded from your PC to the DV-III Ultra,
you will be able to run tests by following the instructions in this section.
S
Yield Programs Menu
Yield test operations are accessed by pressing the "4" key when in the PROGRAM MODES menu.
The following screen is immediately displayed:
YIELD TEST
#5: Sample Tst
ENTER = Run Test
PRINT = Print Test
Figure IV-69
The program number and the program name (in this case, “#5: Sample Tst”) are displayed on the
second line. The program name flashes to indicate that it is the currently chosen test.
Pressing the OPTIONS/TAB key scrolls through the program numbers (0-9) and displays the
corresponding program name.