Viscometer Will Not Return To Zero

Rheometer is not level

• Adjust the laboratory stand

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

Manually lift the spindle nut and rotate counter clockwise until the display shows

20% - 30% deflection.

Release the spindle nut.

The % display should return to 0.0%

± 0.1%.

• Perform calibration check
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

* This procedure does not apply to instruments with ball bearing suspension (See Section I.7).

Display Reading Will Not Stabilize

Special characteristic of sample fluid. There is no problem with the rheometer.

• Refer to Appendix C

Check for erratic spindle rotation

• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Temperature fluctuation in sample fluid

• Use temperature bath control.

No Recorder Response

Be sure the rheometer is not at ZERO reading.

Be sure the recorder is ON and not on STANDBY.

Verify the range settings.

Check cable leads for clean connection.

Recorder Pen Moves In Wrong Direction

Output polarity reversed

• Reverse leads