Installation Guide

Chapter 4 Install the system

35

14.

a. Unplug UPS 1 (in the equipment closet) from AC power
b. Confirm that the Acuity LT Central Station display shows

patient data with no yellow alert messages.
The UPS might emit an audible alarm.

c.

Plug UPS 1 into AC power.

Pass

Fail

14a.

For systems with 4-7 APs only:
a. Unplug the additional UPS (in the equipment closet) from AC

power

b. Confirm that the Acuity LT Central Station display shows

patient data with no yellow alert messages.
The UPS might emit an audible alarm.

c.

Plug in the UPS to AC power.

Pass

Fail

15.

a. Unplug UPS 2 (at the nurses station) from AC power
b. Confirm that the Acuity LT Central Station display shows

patient data with no yellow alert messages.
The UPS might emit an audible alarm.

c.

Plug in UPS 2 to AC power.

Pass

Fail

16.

a. To check the modem, call Welch Allyn Technical Support

(1-800-289-2501 or +1 503-530-7500).

b. Provide the telephone number of the modem, and ask the

technician to check the modem operation.

Pass

Fail

17.

Measure the RSSI for your facility as described on

“Received

signal strength indication (RSSI) check”

on page 36 and complete

the

“Your facility’s RSSI measurement chart”

on page 39.

Confirm that the RSSI for all locations is 50 or less.

This step may require 1-2 hours to complete, depending on your
facility.

Pass

Fail

Test performed by ______________________________________________________ Date _____________

Facility
_____________________________________________________________________________________

Table 5. Final system check for wireless Acuity LT System (continued)

Step

Activity

Pass/fail
(circle
one)

Comments