Chapter 4 Install the system

Welch Allyn Wireless Acuity LT Central Monitoring System

4.

For one of the following patient monitors, take these steps to set the monitor to the
alternate patient simulation mode:

Micropaq monitor:

Press

to

highlight

TOGGLE DEMO MODE, and then press

.

Propaq LTR monitor:

a.

Highlight SpO2, and then press

.

b.

Highlight Setup, and then press

.

c.

Highlight Timings, and then scroll to highlight Demo Mode.

d.

Highlight Low, and then press

to return to the main display.

Wireless Propaq CS monitor:

a.

Press SETUP, WAVE SELECT, and then INSERVICE.

b.

Confirm that the Waveform Window border changes from red (patient alarm
state) to yellow (equipment alert state).

5.

Click in the waveform area of the Waveform Window to print data from this simulated
patient.

Confirm that within two minutes, the printer prints the data. If it does not, verify that
the printer is online and properly connected.

Print area