Chapter 4 Install the system
27
To simulate a shutdown
1.
Unplug both UPS units from AC power.
The UPS units might emit alarm tones.
2.
Confirm that the system continues to operate without disruption, and then plug in
both UPS units to AC power.
3.
Apply power to each of the wireless monitors that will be used with the Acuity LT
System (a maximum of 12 active monitors at one time, or 28 active monitors for a 28-
patient system).
Confirm that within one minute, the Acuity LT Central Station display shows a
waveform window for each patient monitor.
4.
Take the following steps to disconnect all of the monitors from the Acuity LT System:
•
Micropaq monitor:
a.
Press
to highlight ACUITY, and then press
.
b.
Press
to highlight END TELE, and then press
.
c.
When the SAFE TO REMOVE BATTERY message is displayed, remove the
battery.
•
Propaq LTR monitor:
a.
Press
.
b.
Highlight Delete & Shutdown, and then press
.
•
Wireless Propaq CS monitor:
a.
From the Main Menu, press SETUP, ACUITY, and then END TELE.
b.
Turn off power.
5.
Shut down the Acuity LT System as follows:
a.
On the Acuity LT Central Station screen, move the pointer into the gray area in the
upper right hand corner of the screen.
b.
Hold down the
◆
key on the keyboard, and press the right mouse/trackball key.
c.
When the System Menu appears in the upper right corner, click
Shutdown Workstation, and then click Confirm.
d.
After the system shuts down, turn off system power to all Acuity LT System
components and disassemble the system.
If any part of the system is not operating properly, refer to