Installation Guide

Chapter 4 Install the system

33

6.

a. At the Acuity LT Central Station display, click in the waveform

area of the Waveform Window.

b. Confirm that within two minutes, the printer prints simulated

patient data.

Pass

Fail

7.

a. In the Waveform Window, click the status icon.
b. Confirm that an ID Not Confirmed message is shown in the

yellow border of the Virtual Monitor.

Pass

Fail

Table 5. Final system check for wireless Acuity LT System (continued)

Step

Activity

Pass/fail
(circle
one)

Comments