9
93
Care and maintenance
Avoid electrostatic discharge
When humidity in the working environment decreases, the human body and other
insulators can become charged with static electricity due to friction.
To prevent unwanted electrostatic discharge (ESD), follow these standard guidelines:
•
Maintain the recommended humidity of 40% to 60% in the work environment.
•
Dissipate electrostatic charge before performing routine operator maintenance.
Inspect and clean the monitor and accessories
Before cleaning, thoroughly inspect the monitor and all accessories for any signs of
damage, cracks, or improper mechanical function of keypads, switches, connectors, and
printer paper door. While gently bending and flexing cables and tubing, inspect for
damage, cracks, cuts, abrasions, extreme wear, exposed wires or bent connectors.
Confirm connectors securely engage. Report damage or improper function to your service
department.
Equipment
Cleaning instructions
Approved cleaning solutions
Propaq CS Monitor
a
b
a.
Do not use these cleaning solutions (they may damage the monitor): Butyl alcohol, Denatured ethanol, Freon, Mild chlorine
bleach solution, Isopropyl alcohol, Trichloroethane, Trichloroethylene, Acetone, Vesphene II, Enviroquat, Staphene, Misty,
Glutaraldehyde, Fantastik, Formula 409, Cidex.
b.
The monitor may be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood and other body
fluids. (Federal OSHA Standard on bloodborne pathogens: 29 CFR 1910.1030, 12/6/91.) Wex-cide (Wexford Labs, Inc., Kirkwood, MO)
and T.B.Q. (Calgon Vestal Lab., Calgon Corp., St. Louis, MO) are disinfectants that meet OSHA requirements, and are EPA approved.
Wipe away disinfectants with a water-dampened cloth after the manufacturer’s recommended period of time.
Wipe with a nearly-dry cloth moistened
with cleaning solution.
Thoroughly wipe off any excess cleaning
solution. Do not let water or cleaning
solution run into connector openings or
crevices.
c
c.
If liquid gets into the right side panel connectors, it will drain out. If moisture gets into a left side panel connector, dry the connector
with warm air, then check the monitoring functions for proper operation.
Warm water
Liquid soap
T.B.Q.
b
Hydrogen peroxide solution
Coverage
Wex-cide
b
Windex
Ovation
NIBP cuff
Consult manufacturer’s instructions.
Consult manufacturer’s instructions.
Cables, tubing, CO
2
sensor
d
Wipe gently with cloth dampened with
cleaning solution. Do not immerse the
CO
2
sensor in liquid.
Mild detergent solution; also consult manufacturer’s
instructions.
SpO
2
cables
Consult manufacturer’s instructions.
Consult manufacturer’s instructions.
Other accessories
Consult manufacturer’s instructions.
Consult manufacturer’s instructions.