Real-time ecg analog/defib sync – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual
Page 105

Reference Guide
Specifications
101
Real-time ECG analog/defib sync
Special cables are required to interface the defib sync connector to the Physio-Control
LIFEPAK 5 or LIFEPAK 6s defibrillator.
The sync and real-time ECG outputs do not operate during In-service mode.
Signal
Specification
Sync Output
0 to 5V pulse, 100 ± 5ms wide, starts within 35 ms after peak of R-wave. 15 mA short
circuit current.
Real-time ECG Output
Range = ± 6V minimum, centered about 0V, Gain = 1000X, noninverting for lead II,
inverting for all other leads, delay <3 msec, 0.05-100 Hz, going to -5.9V ± 5% during
ECG lead fail. V lead has no Real-Time analog output.
Marker Input (Defib Sync
only)
Normally 0V in, a pulse either + or -3 to ± 15V for 10-70 ms puts a marker in ECG
trace. ~ 5 k
Ω input res.
Shield
Common terminal for other signals
N/C
N/C
GROUND
MARKER IN
SYNC
OUT
N/C
Real-time ECG Output connector
(side-panel view)
Defib Sync connector
(side-panel view)
ECG
N/C
GROUND
SHIELD
N/C
SHIELD