Monitor (environmental) – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual
Page 117

Reference Guide
Specifications
113
Monitor (environmental)
Caution The monitor may not meet performance specifications if it is not used
or stored within these environmental specifications.
Characteristic
Specification
Operating Temperature
0° to 40° C
Shipping and Storage Temperature
-20° to 60° C
Operating Altitude
-2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude
-2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity
15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shipping and Storage Relative
Humidity
15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shock
50 g
Vibration, Random
0.02g
2
/Hz from 10 to 500 Hz, ramping down to 0.002g
2
/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA DO-
160C, Category C.
Vibration, Sinusoidal
0.10 peak to peak inches 5 to 17 Hz, sloping to 0.01 peak to peak inches at
55 Hz, then sloping to 0.0001 peak to peak inches at 2000 Hz. Operating 1
hour per axis, 3 hours per test. Designed to meet RTCA DO-160C, Category
N.
Degree of protection against ingress
for monitors without CO
2
or printer
options
IPX1 rating, drip proof per EN60529: 1991