In-service simulated values – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual
Page 124

120
Specifications
Welch Allyn Propaq Encore Vital Signs Monitor
In-service simulated values
Channel
Display
Initial value
Alternate value
ECG
Waveform
Normal sinus rhythm, 1mV, Lead II
Normal sinus rhythm
ECG
Heart Rate
80 beats per minute
125 beats per minute
RESP
Respiration Rate
12 breaths/minute
31 breaths/minute
P1
Waveform
Arterial
Same as Initial Value
P1
Pulse Rate
80 pulses per min
125 pulses per min
P1
Systolic
121 mmHg
120 mmHg
P1
Diastolic
79 mmHg
85 mmHg
P1
Mean
96 mmHg
103 mmHg
P2
Waveform
Pulmonary Artery
Same as Initial Value
P2
Pulse Rate
80 pulses per min
125 pulses per min
P2
Systolic
25 mmHg
25 mmHg
P2
Diastolic
9 mmHg
12 mmHg
P2
Mean
15 mmHg
18 mmHg
NIBP
Mode
Manual (Auto cancels in-service)
Same as Initial Value
NIBP
Numerics
Actual values from patient
Actual values from patient
T1
Numeric
37.0° C
39.1° C
T2
Numeric
36.4° C
37.4° C
ΔT
Numeric
0.6° C
1.7° C
SpO
2
Waveform
Normal, 2x
Same as Initial Value
SpO
2
Rate
80 pulses per min
125 pulses per min
SpO
2
Numeric
97%
88%
CO
2
Waveform
Normal
Hyperventilating
CO
2
ETCO
2
Numeric
38 mmHg
60 mmHg
CO
2
INCO
2
Numeric
0 mmHg
8 mmHg
CO
2
Breath Rate
12 breaths/minute
31 breaths/minute