Monitor setup – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual
Page 24
20
Getting started
Welch Allyn Propaq Encore Vital Signs Monitor
Monitor setup
Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.
STATSCALE
Automatically readjusts all waveform scales.
ALARMS
Allows access to the Alarms menu.
WAVE SEL
Allows you to turn on and off desired waveforms or NIBP numerics
for display.
TRENDS
Allows access to the Trend settings and display.
MORE
Displays the next setup menu and the following status window:
NEXT
Selects the next setting in the status window.
CHANGE
Changes the currently selected display setting. (Pressing CHANGE
at PATIENT MODE allows you to choose between Adult, Pediatric,
and Neonatal in a Patient Mode window.)
PRINTER
Allows access to the Printer Menu.
MORE
Allows access to the Time/Day window.
CURRENT
SOURCE
When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It will always be CO
2
if CO
2
is active.
Otherwise, it will be ECG/RESP.
SELECTED
SOURCE
The user-selected HR/PR source is displayed along with the HR/PR
source currently being used by the monitor.
SWEEP (mm/s) The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/
sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.
ALARM TONE
Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.
STATSCALE
MORE
WAVE SEL
TRENDS
ALARMS
mmHg
NEXT
MORE
WAVE SEL PRINTER
CHANGE
MCO
2
BR
SpO
2
PA
SETUP
HR/PR
HR/PR RR/BR
CURRENT SOURCE : ECG MCO2
SELECTED SOURCE : ART ---
SWEEP (mm/s) : 12.5 3.13
ALARM TONE : HIGH
HR/PR TONE : LOW
PATIENT MODE :
ADULT
35/ 18
( 85)
35 12
92