beautypg.com

Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual

Page 72

background image

68

Patient monitoring

Welch Allyn Propaq Encore Vital Signs Monitor

the monitor to within this range automatically cancels this message and restores
operation.

ALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is
changing at a rate greater than 100 mmHg/minute. When the rate of change is back
within the 100 mmHg/minute range, disconnect and reconnect the CO

2

sensor to the

Propaq.

ALTIMETER NOT CALIBRATED - EEPROM - The sidestream CO

2

option has not been

calibrated. Refer the Propaq to a Biomedical Engineer for calibration.

AMBIENT TEMPERATURE TOO HIGH. The sensor temperature is too high. The
sidestream CO

2

option is disabled until the ambient temperature is within the

operating range specifications.

AMBIENT TEMPERATURE TOO LOW. The sensor temperature is too low. The
sidestream CO

2

option is disabled until the ambient temperature is within the

operating range specifications.

CALIBRATION ERROR, SERVICE REQUIRED. Send the Propaq to a Biomedical
Engineer for service.

DEGRADED WAVEFORM, SERVICE REQUIRED. Send the Propaq to a Biomedical
Engineer for service.

LACK OF WAVEFORM, SERVICE REQUIRED. Send the Propaq to a Biomedical
Engineer for service.

MOTOR FAILURE, SERVICE REQUIRED. The sensor hardware has failed. Send the
Propaq to a Biomedical Engineer for service.

NO WATERTRAP DETECTED. There is no sidestream CO

2

watertrap or CO

2

watertrap installed. Install a watertrap or CO

2

watertrap adapter.

OCCLUSION - CHECK EXHAUST PORT/TUBING. Blockage has been detected on the
pneumatic exhaust port. Check the exhaust port and related tubing for occlusions.
Make sure that the sampling line and any inputs to the patient breathing apparatus are
not connected to the exhaust port.

OCCLUSION - CHECK WATERTRAP/TUBING. Blockage has been detected on the
sidestream CO

2

input. Check the watertrap, sample line, and any connected tubing

for occlusion.

PUMP FAILURE, SERVICE REQUIRED. The pump is not able to maintain the target
flow rate. Send the Propaq to a Biomedical Engineer for service.

SIDESTREAM STICK EEPROM FAILURE. Send the Propaq to a Biomedical Engineer
for service.

SSP BOARD EEPROM FAILURE. Send the Propaq to a Biomedical Engineer for
service.