HR/PR TONE
Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.
PATIENT MODE Pressing CHANGE in this selection displays the following Patient
Mode window:
If you press NEONATAL, PED, or ADULT, a confirmation window
appears, requiring you to confirm your selection:
Note
Whenever you change the patient mode, the alarm limit settings are
automatically changed to the defaults for that mode. If Custom
settings have been set for that mode, the defaults are the Custom
mode settings. If no Custom settings have been set, the defaults are
the Factory Mode settings. See page 23 for more information about
patient modes.
If you change the patient mode, the CO
2
alarm limits in the new
mode might vary slightly from the originally-programmed CO
2
limits
for the new mode. Check the CO
2
alarm limits.
If you press SETUP in the previous Patient Mode window, the Mode
Setup window appears. This allows you to set custom patient
modes and powerup defaults as described on page 23.
SETUP
BATTERY: 9.2 VOLTS
mmHg
NEONATAL
PREV MENU
ADULT
PED
MCO
2
BR
SpO
2
PA
PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:
NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT: > 9 YEARS
35/ 18
( 85)
35 12
92
mmHg
YES
NO
MCO
2
BR
SpO
2
PA
PATIENT MODE
CHANGING SETTINGS AND MODE
35/ 18
( 85)
35 12
92
ARE YOU SURE?
TO:
ADULT ALARM LIMITS WILL BE SET
CUSTOM: ADULT