Getting started

Welch Allyn Propaq Encore Vital Signs Monitor

You can program the Propaq Encore to power up in any of the Factory patient modes or
the Custom patient modes. You can also change the current patient mode during
operation.

From the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window
appears:

The asterisk (*) indicates which patient mode is currently selected for powerup.

Note

The alarm for apnea cannot be turned off at any time.

Note

Whenever you change the patient mode, the alarm limit settings automatically
change to the settings for that mode.

NEXT

Selects the next setting in the status window.

POWERUP* Selects the highlighted patient mode (and its associated settings) as

the powerup mode. The selected powerup mode is marked by an
asterisk (*). (This does not change the current patient mode.)

USE NOW

Selects the highlighted patient mode (and its associated settings) as
the current patient mode. (This does not affect the powerup mode.)

SAVE

Use this button to reprogram the settings of Custom patient mode:

1.

Make sure the patient mode you want to reprogram (ADULT, PED
or NEO) is currently used. (To change patient modes, highlight the
desired mode and press USE NOW, YES.)

2.

Exit the Mode Setup window, then use other menus and buttons to
set the monitor settings and alarm limits as desired.

3.

Re-enter the Mode Setup window, highlight the desired Custom
mode, and press SAVE, YES.

BATTERY: 9.3 VOLTS

mmHg

NEXT

PREV MENU

USE NOW

POWERUP*

MCO

2

BR

MODE SETUP

FACTORY:

NO ZERO

38 12

97

*ADULT

PED

NEO

CUSTOM

ADULT

PED

NEO

* = PATIENT MODE ON POWERUP

SAVE

P

2

:

Sp02

ZEROED