Factory default settings, Factory default, Settings – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual
Page 122

118
Specifications
Welch Allyn Propaq Encore Vital Signs Monitor
Factory default settings
Characteristic
Specification
Date
a
MO/DA/YR, DA.MO.YR, or YR/MO/DA
Decimal
a
. (Period)
HR/PR Sweep
25 mm/sec
RR/BR Sweep
6.25 mm/sec
Alarm Tone
MEDIUM
HR/PR TONE
LOW
HR/PR SOURCE
ECG
RR/BR Source
CO
2
if ON or ECG (if CO
2
OFF)
Patient Mode
Adult
ECG Bandwidth
Monitor
ECG Size
1 mV/cm
ECG Lead
II
ECG Filter
a
60 Hz
ECG Pacer
ON
RESP size
2X
RESP lead
Ld2
RESP sweep
6.25 mm/sec
RESP ON/OFF
ON
RESP window
ON
IBP Range
0 to 180 mmHg
IBP Rescale
0 to 140 mmHg
IBP Mode
RESCALE
Invasive Pressure Formats
Label dependent
NIBP Mode
MANUAL
NIBP Auto Time
15 min
SpO
2
SIZE
2x
SpO
2
C-LOCK
OFF
SpO
2
Response
NORMAL
TEMP F/C
a
Celsius
CO
2
Range
0 to 60 mmHg
CO
2
Sweep
6.25 mm/sec
CO
2
Response
NORMAL
CO
2
Units
a
mmHg
CO
2
Gas Compensation
OFF
Sidestream CO
2
Flow Rate
Adult: 90 ml/minute
Ped: 90 ml/minute
Neonate: 90 ml/minute
(The flow rate cannot be programmed to a different value in a Custom Patient Mode.
See Custom Patient Modes.)
Display Wave Select
All waves are on except NIBP
Trend Group
NIBP
Alarm Limits
All are ON except P2
HR Limits
Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic
Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean
Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic
Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg