Multiplexor priority setting – ADS Environmental Services Model 3600 530002 A2 User Manual
Page 94
5-20
ADS 3600 Flow Monitor O&M Manual
J5
1
2
3
J5
1
2
3
J5
1
2
3
Backup Battery
O perating
Battery in
Storage
Three-position
Jum per Header
w ithout Jum per
To disable the backup battery, locate Jumper 5 (a three position jumper header). Move the jumper to cover positions 2
and 3. The battery is now disabled. When the jumper is covering positions 1 and 2, the backup battery is operating.
Multiplexor Priority Setting
The serial multiplexor electronics are located inside the EMU. On the multiplexor
power electronics printed circuit board (PCB) is a three-position jumper that can be
used to specify which port will have priority, the direct connect port or the modem
port. The 3600 monitor does not support the direct connect port. Therefore, always
maintain the jumper in the modem position. This jumper is labeled “J1”.