Page 72: Drill and mount the anchor bolt to the pipe wall
Ring, Sensor, and Special Installations
4-39
l
This mounting method uses a special flat strip of rigid aluminum instead of the
standard band. It is stronger and is less likely to become damaged during high
flow conditions.
l
It is imperative that the ring (below the water's surface) be secured with an
anchor bolt and placed within 25 to 75 mm (1 to 3 in.) of the velocity sensor.
This will prevent the velocity sensor from becoming clogged should debris
accumulate between the pipe and the metal band.
A.
B.
C.
D.
Point F.
Crown
Anchor Bolt
(1 of 5)
Flow line
1/2 Ring
93012
Velocity Sensor
Silt
E.
1/2-Band Velocity Mount (Special Mount)
Installation
l
Position the sensor portion of the band in the flow, below minimum flow levels
and out of silt.
l
Orient the band in a fashion that allows one of the pre-drilled holes to be right
at the water surface.
l
Spot mark with a drill.
l
Drill and mount the anchor bolt to the pipe wall.
l
Attach the band to the anchor bolt with a washer and a 7/16 in. (11 mm) nut.
l
Position the submerged portion of the band until it is flush with the bottom of
the pipe. A slight downward bend of the band right below the water surface
may aid in ensuring that the band and sensors are flush.